Mars Pro 1200 Class II Biosafety Cabinet
Triple Filter System Microbe Capture Trap
Triple filter system with H14 HEPA unique microbe capture trap, the filter efficiency is high up to 99.9995%.
Introduction
Mars Pro 1200 is built based on the common Class II, Type A2 BSC, with one additional H14 HEPA filter under the operating worktop. Enter the lower filter under the operating worktop for primary filtration, and then pass through the upper HEPA filter for secondary filtration. 2 sets of triple HEPA filter systems can effectively filter 99.99999975% of 0.3μm particles. It can be used for microbiology and virology research, especially on infectious pathogens; stem cell/gene therapy; research on cytotoxic substances, etc.
Features
The main HEPA filter under the worktop has 4 sets of 12 built-in "microbial traps", inside which the pathogenic microorganisms can be captured and well-kept, until later disposal and exchange of filters with no prior decontamination being required;
Air pathways and channels, filters, and fans remain contamination-free.;
Safe to replace filter whilst the cabinet is operating – able to seal and remove the used filters under the worktop in a negative pressure environment.